Commands Listed in Alphabetical Order
MEASUrement:
MEAS<x>:STATE
Sets or returns whether the specied measurement slot is computed and displayed.
The measurement slot is specied by <x>, which ranges from 1 through 4.
For a measurement to display, you must have selected a source waveform
and dened the measurement you want to take and display. You select the
measuremen
tusingtheMEASUrement:MEAS<x>:SOURCE[1] command.
You dene the measurement type using the MEASUrement:MEAS<x>:TYPe
command.
Group
Measurement
Syntax
MEASUrement:MEAS<x>:STATE {OFF|ON|<NR1 >}
MEASUrement:MEAS<x>:STATE ?
Related Commands
MEASUrement:MEAS<x>:SOURCE[1], MEASUrement:MEAS<x>:TYPe
Arguments
OFF disables calculation and display of the specied measureme nt slot.
ON enables calculation and display of the specied measurement slot.
<NR1> = 0 disables calculation and display of the specied measurement slot; any
other value enables calculation and display of the specied measurement slot.
Examples
MEASUREMENT:MEAS2:STATE ON computes and d isplays the measurement
dened as mea surement 2.
MEASUREMENT:MEAS1:STATE? might return :MEASUREMENT:MEAS1:STATE 0
indicating that measurement dened for measurement slot 1 is disabled.
MEASUrement:MEAS<x>:STDdev? (Query Only)
Returns the standard deviation of values accumulated for this measurement since
the last statistical reset. Measurements are specied by <x>, the measurement
slots, from 1 through 4.
Group
Measurement
Syntax
MEASUrement:MEAS<x>:STDde v?
2-222 MSO4000 and DPO4000 Series Programmer Manual