Commands Listed in Alphabetical Order
Syntax
MATH[1]:VERTi
cal:POSition <NR3>
MATH[1]:VERTical:POSition?
Related Commands
CH<x>:POSition, REF<x>:VERTical:POSition
Arguments
<NR3> is the desired position in divisions from the center graticule.
Examples
MATH1:VERTICAL:POSITION 1.3E+00 positions the Math 1 input signal
1.3 divisions higher than a position of 0.
MATH1:VERTICAL:POSITION? might return :MATH1:VERTICAL:POSI TION
-1.3000E+00
indicating that the current position of Math 1 is 1.3 divisions
below the center graticule.
MATH[1]:VERTical:SCAle
Sets or returns the vertical scale of the currently selected math type.
Group
Math
Syntax
MATH[1]:VERTical:SCAle <NR3>
MATH[1]:VERTical:SCAle?
Related Commands
CH<x>:SCAle, REF<x>:VERTical:SCAle
Arguments
<NR3> is the scale-per-division in the current math vertical units. The range is
from 1 .0E-12 through 500.0E+12.
Examples
MATH1:VERTICAL:SCALE 100E-03 sets the Math scale to 100 mV per division.
MATH:VERTICAL:SCALE? might return :MATH:VERTICA L:SCALE
1.0000E+00
indicating that the current scale setting o f Math is 1 V per division.
MATH[1]:VERTical:UNIts
Returns the math waveform vertical measurement unit value.
Group
Math
MSO4000 and DPO4000 Series Programmer Manual 2-203