Performance Verification
P5205 Instructi on Manual
29
Vise
Pliers
Modified BNC adapter
Plastic cover removed
Figure 12: Removing plastic covers from bi nding posts
Setup
WARNING. These procedures require the application of high voltage
to the inputs of the P5205 probe. Because this adapter has exposed
metal surface, only qualified personnel should perform testing w ith
voltage levels that exceed 30 V rms. All pertinent safe ty rules and
guidelines for ele vated voltage measurements should be followed and
adhered to.
Connect the output of the probe to the Channel 3 input of the test
oscilloscope. Let the probe warm up for 20 minutes.