Tektronix TDS500C Instrukcja Użytkownika Strona 99

  • Pobierz
  • Dodaj do moich podręczników
  • Drukuj
  • Strona
    / 148
  • Spis treści
  • BOOKMARKI
  • Oceniono. / 5. Na podstawie oceny klientów
Przeglądanie stron 98
Performance Tests
TDS 500C, TDS 600B & TDS 700C Performance Verification and Specifications
1–83
Figure 1–40: Sync duty cycle test: critically adjusted pulse
H Press TRIGGER MENU.
H Press the main-menu Type pop-up until you select Video.
If the TRIG’D LED is not lighted, check that the CH1 – Width and
CH1 Period measurements are adjusted correctly. See note above.
CONFIRM that the setup is correct and the oscilloscope will trigger.
H CONFIRM that the TRIG’D LED is lighted and the waveform is
stable.
H Disconnect the signal source from CH1, wait a few seconds, then
reconnect the signal.
H CONFIRM that the TRIG’D LED is lighted and the waveform is
stable.
H Press Sync Polarity.
H Press Pos Sync.
H Push the pulse generator COMPLEMENT button out.
H CONFIRM that the TRIG’D LED is lighted and the waveform is
stable.