Reference
TDS1000/2000-Series Digital Oscilloscope User Manual
77
When there is enough waveform noise, a typical peak detect display
shows large black areas. The TDS1000- a nd TDS2000-series
oscilloscopes display this area with diagonal lines to improve display
performance.
Typical peak detect display
TDS1000/TDS2000 peak detect display
Average. Use Average acquisition mode to reduce random or
uncorrelated noise i n the signal you want to display. Data is acquired
in sample mode, then a number of waveforms are averaged together.
Select the number of acquisit ions (4, 16, 64, or 128) to average for
the waveform.
RUN/STOP Button. Push the RUN/STOP button when you want the
oscilloscope to continuously acquire waveforms. Push the button
again to stop the acquisition.
SINGLE SEQ B utton. Push the SINGLE SEQ button when you want the
oscilloscope to acquire a single waveform and then stop. Each time
you push the SINGLE SEQ button, the oscilloscope begins to acquire
another waveform. After the oscilloscope detects a trigger it
completes the acquisition and stops.