Tektronix TDS500C Instrukcja Użytkownika Strona 61

  • Pobierz
  • Dodaj do moich podręczników
  • Drukuj
  • Strona
    / 148
  • Spis treści
  • BOOKMARKI
  • Oceniono. / 5. Na podstawie oceny klientów
Przeglądanie stron 60
Performance Tests
TDS 500C, TDS 600B & TDS 700C Performance Verification and Specifications
1–45
Check long–term sample
rates and delay time
accuracies against limits.
4
Align the trigger T to the
center graticule line.
1
Set
horizontal
mode.
2
Set horizontal scale
and delayed time.
3
Figure 1–13: Measurement of accuracy — Long-term and delay time
3. Disconnect the hookup: Disconnect the cable from the generator output at
the input connector of CH 1.
Trigger System Checks
These procedures check those characteristics that relate to the Main and Delayed
trigger systems and are listed as checked in Specifications.
Equipment
required
One medium-frequency sine wave generator
(Item 19 or, optionally, item 25)
One 10X attenuator (Item 1)
One 50 , precision coaxial cable (Item 5)
Prerequisites See page 1–15
Check Accuracy (Time) for
Pulse-Glitch or
Pulse-Width Triggering
Przeglądanie stron 60