Syntax and Commands
AFG3000 Series Arbitrary/Function Generators Reference Manual 3-87
[SOURce[1|2]]:VOLTage:LIMit:LOW
This command sets or queries the lower limit of the output amplitude low level for
the specified channel. If your instrument is a dual-channel model and the
[SOURce[1|2]]:VOLTage:CONCurrent[:STATe] command is set to ON, then the
low level lower limit of the other channel is the same value.
Group
Source
Related Commands
[SOURce[1|2]]:VOLTage:CONCurrent[:STATe]
Syntax
[SOURce[1|2]]:VOLTage:LIMit:LOW {<voltage>|MINimum|MAXimum}
[SOURce[1|2]]:VOLTage:LIMit:LOW? {MINimum|MAXimum}
Arguments
<voltage>::=<NRf>[<units>]
where
<NRf> is the lower limit of output amplitude.
<units>::=[mV | V]
Returns
<voltage>
Examples
SOURce1:VOLTage:LIMit:LOW 10mV
sets the lower limit of CH 1 output amplitude to 10 mV.