Tektronix AWG610 Instrukcja Użytkownika Strona 417

  • Pobierz
  • Dodaj do moich podręczników
  • Drukuj
  • Strona
    / 466
  • Spis treści
  • BOOKMARKI
  • Oceniono. / 5. Na podstawie oceny klientów
Przeglądanie stron 416
Appendix B: Performance Verification
AWG610 Arbitrary Waveform Generator User Manual

3. Follow the substeps below to set the AWG610 Arbitrary Waveform
Generator controls:
a. Push UTILITY (front-panel)System (bottom)Factory Reset
(side)OK (side).
b. Load the MODE.WFM file.
Refer to Loading Files on page B–10 for file loading procedures.
c. Push Horizontal Menu (front panel)CLOCK (side).
d. Push 2, 0, 0, and M (SHIFT+7) keys in this order or used the general
purpose knob to set the internal clock frequency to 200 MHz.
4. Push the RUN button.
The LED above the RUN button is on.
5. Verify that the 10 MHz REF OUT clock signal amplitude is equal to or
larger than 1.0 V, and the clock signal period is about 100 ns.
6. Disconnect and remove the BNC cable.